MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 2 of 12
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
5 - 2
Mechanical Checks
1.
Mount the side covers (1/Fig. 2) with six screws (2/Fig. 2) on each side. Start with
the right side.
Front cover
5
1.
The side covers must be flush at the front. If necessary, loosen the screws and
adjust the side covers.
2.
Carefully fit the front cover (4/Fig. 2) so that both openings engage with the lugs of
the side covers.
3.
Carefully swing the front cover upwards and let it snap in position over the side
covers.
4.
Fasten the front cover to the side covers with two short screws (5/Fig. 2).
Do not forget to fit the sign (3/Fig. 2) when mounting the side cov-
ers. Use the two longer screws.
Make sure that the longer screws are not used at the back of the
covers at the upper location (6/Fig. 2) since this may damage the
board underneath the cover.
Risk of damages.
If the covers are exposed to internal stress, cracks might arise.
The following work must be carried out with caution.
The front cover must not press against the collimator.
Risk of damage.
Be careful while attaching the front cover.
NOTE
NOTE
CAUTION
CAUTION