Power up the MAMMOMAT Stand
Measuring the line resistance
3
1.
Connect the line resistance meter to the mains input terminal on the stand.
2.
Mains supply
ON
.
3.
Carry out measurement.
4.
Mains supply
OFF
.
To achieve the maximum output, the line resistance must not exceed the following values:
Checking the supply voltages
3
The correct supply vo5 V, +15 V and +28 V have already been tested in the fac-
tory. Only a visual check of the function is therefore needed.
1.
Switch the system
ON
.
2.
Check that the following LED indicators (top to bottom) on the mains input
converter (generator) are on:
Nominal Input Voltage
[VAC]
Line Resistance Ri
max
[
c
]
208
0,45
230
0,50
240
0,60
277
0,65
400 (2-phase)
0,85
Tab. 1
Measuring the line resistance
LED check
28 VP
5 VF
- 15 V
+ 15 V
28 VS
5 V
Tab. 2
LED check