6 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 4
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Mains connection and power supply
6
Checks before powering up the generator
6
On P.C. board D702:
1.
Set switch S1(UZW) to "OFF" (lower position).
2.
Set switch S2 (SS) to "OFF" (lower position)
.
Measuring the line resistance
6
1.
Remove fuses F10 (1) and F20 (2) or, when single-phase supply, fuse F10 (1).
2.
Connect line resistance meter to fuse holders F10 and F20.
3.
Mains supply "ON".
4.
Carry out measurement.
5.
Mains supply "OFF".