Start-Up of Multiline System
4 - 7
Siemens AG
RXL2-120.815.01
Page 7 of 8
LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
Shock Wave Pressure Test
4
•
Move the shock wave head into the coupling position.
•
Apply ultrasound contact gel to the coupling bellows.
•
Attach and secure PTU adapter M with screws.
•
Selection of energy step and measured values, see Service Instructions Shock Wave
Pressure and Position Control SPL2-120.074.01.
•
Release shock waves.
•
Remove PTU adapter M and reinsert the rubber plugs.
•
Move the shock wave head into the park position after completing the measurement.