A.5 Flowrate Calibration and Calibration Tables
FUG1010 IP65 NEMA 4X
188
Operating Instructions, 01/2013, A5E02951519-AC
Some applications may require an output adjustment to match an official external reference.
The [Calibrate Flowrate] menu allows you to select a calibration mode. The right-hand
column shows the active calibration mode. You can select Intrinsic (factory) and Kc (Slope
Correction) Calibration. Selecting either of the external calibration modes will not eliminate
the Intrinsic (factory) calibration. You can use this menu cell to switch between Intrinsic and
Kc at any time.
Kc Calibration
For most applications, the measured flow range produces a linear meter response.
Therefore, the Kc (slope correction) calibration is the preferred method since it only requires
a single correction factor for all the flow rates encountered.
Note
Changing the calibration can cause profound changes in flow meter operating
characteristics. Use only the most respected flow standard to obtain a correction factor. The
percentage entered must provide an accurate and consistent shift across the entire flow
range anticipated for the application.
Kc Factor
To obtain the Kc factor, compare flow total data taken simultaneously from the flow meter
and a reference meter whose accuracy meets the required standard. Allow both meters to
accumulate flow total data long enough to average out any differences due to flow fluctuation
between the two meter locations. Compare outputs of the two totalizers to determine
percentage increase (+) or decrease (-) that is necessary to produce the best average
correlation between the flow meter and the reference standard.
Selecting the Kc Factor
When the [Kc] menu cell is selected, the flow meter imposes this percentage slope
adjustment of its rate output. Output data is zeroed and corrected for the Reynolds number
(flow profile compensated), however, a percent change in the rate output is imposed based
on the data entered in this cell. The number entered by the user is evaluated into a slope
correction factor by dividing it by 100 and algebraically adding it to 1. The resulting factor is
used as a multiplier on the rate register of the instrument. Thus an entry of -3% will multiply
the rate register by 0.97, for example.
To calculate Kc:
.F
$FWXDO5DWH
,QGLFDWHG5DWH
[
Содержание FUG1010 IP65 NEMA 4X
Страница 9: ...Table of contents FUG1010 IP65 NEMA 4X 8 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 23: ...Safety notes 2 6 Certificates FUG1010 IP65 NEMA 4X 22 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 93: ...Commissioning 6 3 Installation Menus FUG1010 IP65 NEMA 4X 92 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 143: ...Service and maintenance 9 4 Battery disposal FUG1010 IP65 NEMA 4X 142 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 165: ...Troubleshooting 10 5 Force Transmit FUG1010 IP65 NEMA 4X 164 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......
Страница 219: ......
Страница 220: ......
Страница 221: ......
Страница 222: ......
Страница 223: ......
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 227: ......
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ......
Страница 235: ......
Страница 236: ......
Страница 237: ...21614 C 1011NFPS 7 ...
Страница 238: ...1011NFPS 7 C 21614 ...
Страница 239: ...1011NFPS 7 C 21614 ...
Страница 240: ...21614 C 1011NFPS 7 ...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...21614 C ...
Страница 247: ......
Страница 248: ......
Страница 249: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 250: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 251: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 252: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 266: ...Glossary FUG1010 IP65 NEMA 4X 200 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 270: ...Index FUG1010 IP65 NEMA 4X 204 Operating Instructions 01 2013 A5E02951519 AC ...