Commissioning
6.3 Commissioning and operation
FIDAMAT 6
62
Operating Instructions, 01/2019, A5E00222135-04
6.3.3
State change due to fault / maintenance request
In the following tables, you will find possible automatic state changes of the device:
Table 6- 1
Change of state due to fault
Fault
Last state
New state
S1: Parameter memory
- - -
S2: Pump does not start up
(FIDAMAT 6-E only)
Warming-up | Standby
Standby
S3: Flame does not ignite
Warming-up
Pause
S4: External fault
- - -
S5: Oven temperature
Warming-up | Standby | Measuring
| Pause
Pause without heating
S7: Flame temperature
Warming-up | Standby | Measuring
| Pause
Pause
S8: Combustion air/combustion
gas pressure
Warming-up | Standby | Measuring
| Pause
Pause
S9: Heating is off
Warming-up
Pause without heating
S10: 24 h RAM/Flash check
- - -
S12: Line voltage supply
Warming-up | Standby | Measuring
| Pause
Pause
S13: Hardware
Warming-up | Standby | Measuring
| Pause
Pause
S14: Measured value > Full-scale
value of characteristic
- - -
S15: Calibration aborted
- - -
Table 6- 2
Change of state due to maintenance request
Maintenance request
Last state
New state
W1: Calibration difference too
great
- - -
W4: Set clock
- - -
W5: Combustion air/combustion
gas pressure
Standby | Measuring
Warming-up
W6: Temperature sensor display
- - -
W7: Electronics/hardware tempera-
ture
- - -
W8: Flame is out
Standby | Measuring
Warming-up
W9: External maintenance request - - -
W10: AUTOCAL check deviation
- - -
After you have eliminated the cause of a maintenance request or fault, exit the current
operating state using function 91.
Содержание FIDAMAT 6
Страница 1: ...FIDAMAT 6 ...
Страница 2: ......
Страница 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 16: ...Introduction 1 9 Information on recycling FIDAMAT 6 14 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 42: ...Installing 4 4 Ventilation of the device FIDAMAT 6 40 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 146: ...Service and maintenance 9 10 Other faults FIDAMAT 6 144 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 186: ......
Страница 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Страница 201: ......
Страница 202: ......