Operation
5.10 Parameters
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-31
Note
Functional check ("CTRL")
If the functional check is active, the measured value might be falsified! The measured
value must therefore be checked – after deactivating the functional check.
As long as you are in code level 1 or 2, in "operating mode" in other words, the functional
check is active. If you parameterized the analog output with "Memory on" under
function 77, this remains in the fault current until you leave the corresponding code level,
and therefore "operating mode" and return to "Coded display mode" (press [MEAS]
twice).
Table 5-5
Status messages
What appears in the display depends on functions 52 and 53
Status
Fct. 53 "□" Fct. 52 "□"
Fct 53 "■"
Fct. 52 "■"
Fct 53 "■"
Calibration: CAL none
CAL
□ CAL
■ CAL; Calibration running (also in
Autocal)
Measured value
memory: STO
none
STO
□ STO
■ STO; Analog output lies on
memory (see function 77)
Limit: LIM
none
LIM
□ LIM
■ LIM; Limit has been exceeded or
fallen short of (see function 51)
Automatic
measuring
range switch:
AR
none
AR
□ AR
■ AR; Measuring range switch
(actively possible), see function 40
Functional
check: CTRL
none
□ CTRL or ■ CTRL
(functional check
cannot be
switched off via fct.
52)
□ CTRL
■ CTRL; Device is in the "Warm-up
phase" operating state or in
"operating mode" or calibration is in
progress.
If an error occurs during operation, the message "Maintenance request", "Fault" or
"Measuring protection" appears in the status line, depending on the severity of the error. This
message is alternately displayed with the status messages.
In addition, when communication goes via the ELAN interface, the message "Remote" is
displayed – also alternating – with the display of the device status.
See also
Select measuring ranges (function 40) (Page 5-23)
Limits (function 51) (Page 5-28)
Analog output memory (function 77) (Page 5-53)