Operation
5.11 Configuration
FIDAMAT 6-G
5-60
Operating instructions, 07/2006, A5E00478463
5.11.17
Linear temperature compensation (function 86)
86 Lin.temp.comp. THC
aftercompensation of
the zeropoint
aftercompensation of
the span
Figure 5-66 Compensating temperature
The FIDAMAT 6 is temperature compensated both in the zero point as well as for the span.
If an additional temperature error occurs during operation, compensate it with this function.
"Recompensation at the zero point":
Assuming an average physics temperature, define two different correction values for ranges
of increased temperature and of decreased temperature.
"Recompensation at the measured value (span)":
The procedure is the same as for the zero, but the change refers to the measured value.
Note
If the zero point deviates negatively when the temperature changes, it carried a positive sign,
and vice-versa. The same applies to the measured value.
5.11.18
Error on/off (function 87)
87 Error On/Off
THC
S1
Paramter memory
S2
S3
Flame does not ignit
S4
External fault
...continue