Performing a zero point adjustment
1. Flush out the flowmeter until a homogenous flow is established and the tubes are completely
filled.
/
Figure 6-4
Best practice zero point adjustment with a by-pass line and two shut-off devices
2. Establish zero flow for example by closing the shut off-valves.
3. Wait 1 to 2 minutes for the system to settle, and then perform zero point adjustment as
described in the following steps.
4. Select "Device" > "Wizard - Zero Point Adjustment..." from the main menu of SIMATIC PDM to
perform an automatic zero point adjustment.
Commissioning
6.3 Remote commissioning with PDM FCT010
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
63