Compact design
SITRANS MASS 2100 (Di 3, 6, 15) is a single mechanical unit where the transmitter is directly
mounted on the sensor.
Remote design
SITRANS MASS 2100 (DI 1.5, 3, 6, 15) or FC300 (DN4) sensor unit is remotely connected to a
SITRANS FCT010 transmitter.
3.4.1
Sensor design
All primary process measurement of mass flow, volume flow, density and medium temperature
are made in the FCT010.
The sensor comprises one single continuous bended tube, with no internal welds. The sensor is
available in an intrinsically safe (IS) design for hazardous area installations.
The sensors are available in AISI 316L stainless steel and Hastelloy C22 or C4. The enclosure is
made of AISI 316L stainless steel.
The sensor enclosure can be equipped with a pressure guard or flushed with dry inert gas at the
threaded ports for non-hazardous applications only.
In the remote configuration, the FCT010 is available in an aluminum enclosure with an ingress
protection grade of IP67/NEMA 4X. For communication and power supply a 4-wire connection
can be made via M12 plug and socket or cable gland/conduit entry for cable termination.
Description
3.4 Design
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
19