Table C-20
Alarms
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
3012
Unsigned / 4 Alarm Group 1
The following bit is set in
case of active alarm:
• Bit 4:
Power Supply Malfunc‐
tion
• Bit 6:
Temperature Circuit Mal‐
function
• Bit 10:
Measurement Out Of
Range
• Bit 14:
Calibration Malfunction
• Bit 15:
Compensation Out Of
Range
• Bit 17:
Pickup Malfunction
• Bit 23:
Driver Malfunction
• Bit 26:
Measurement Out Of
Range
• Bit 27:
Mass Flow Max Limit Ex‐
ceeded
• Bit 28:
Volume Flow Max Limit
Exceeded
• Bit 29:
Density Max Limit Excee‐
ded
• Bit 30:
Min Tube Temp Exceeded
• Bit 31:
Max Tube Temp Exceeded
-
-
Read Only
Modbus communication FCT010
C.4 Modbus holding registers tables
SITRANS MASS 2100 & FC300 (FCT010/DSL)
174