B.2.5
Zero point adjustment
MASS 2100
Sensor size
Default value kg/s
Standard deviation limit
DI 1
0
DI 1.5
0.000002
DI 2.1
0.000002
DI 3
0.00002
DI 6
0.0001
DI 15
0.0004
Sensor size
Default value kg/s
Zero offset limit
DI 1
0.0000272
DI 1.5
0.0000542
DI 2.1
0.000117
DI 3
0.000250
DI 6
0.00139
DI 15
0.0106
Sensor size
Default value kg/s
Default value kg/h
Low flow cut-off
DI 1
0.0000272
0.0980
DI 1.5
0.0000542
0.195
DI 2.1
0.000117
0.42
DI 3
0.000250
0.90
DI 6
0.00139
5
DI 15
0.0106
38
FC300
Sensor type
Sensor size
Default value kg/s
Standard deviation limit
DN4 HP
2
0
DN4 C22
3
0.000002
DN4 316
3.5
0.000002
Technical reference
B.2 Sensor dimension dependent default settings
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
145