Memory card
Capacity is 75 % used
Memory card
Capacity is 100 % used
The table explains the meaning of possible device status and actions for the user or service.
Depending on the type and availability of local display of the device, status is shown as
SIMATIC and/or NAMUR NE107 symbol. Additionally the symbol for the according device
status is visualized in remote engineering, asset management or process control systems.
Refer to the manual of your automation system how diagnostics states are displayed.
Symbols
Table 10-1
Mapping of device states in SIMATIC products (e.g. SIMATIC PDM) and NE107
SIMATIC PDM / PLC
NAMUR NE 107
Description
Measure
Icon
Device
status
Pri‐
ority
Icon
Device
status
Pri‐
ority
Mainte‐
nance
alarm
1
Failure
1
Output signal invalid due to malfunc‐
tion in the field device or its peripher‐
als.
Maintenance is required
immediately.
Mainte‐
nance de‐
manded
2
Mainte‐
nance re‐
quired
3
Output signal is still valid but wear re‐
serve is nearly exhausted, or a func‐
tion will soon be restricted.
Maintenance is strongly
recommended as soon as
possible.
Mainte‐
nance re‐
quired
3
Mainte‐
nance re‐
quired
3
Output signal is still valid. No func‐
tional restriction has been detected
but wear reserve is expected in the
next weeks.
Maintenance of device
should be planned.
Manual
mode
4
Function
check
2
Output signal temporarily invalid (e.g.
frozen) due to ongoing work on the
device.
Disable manual mode via
HMI or engineering sys‐
tem.
Simula‐
tion mode
5
Function
check
2
Output signal temporarily does not
represent the process because out‐
put based on a simulation value.
Disable simulation mode
via HMI or engineering
system or restart device.
Out of
service
6
Function
check
2
Output signal does not represent
process value.
Device mode is set to "Out of service".
Disable “Out of service”
and activate normal opera‐
tion mode.
Configura‐
tion error
7
Failure
1
Output signal invalid due to a param‐
eter setting, an interconnection error
or a configuration error in HW.
Check device hardware
configuration or parameter
settings via HMI or engi‐
neering system.
Diagnosing and troubleshooting
10.1 Device status symbols
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA
187
Содержание FC300
Страница 9: ...Index 515 Table of contents MASS 2100 FC300 FCT030 Operating Instructions 06 2017 A5E39789040 AA 9 ...
Страница 10: ...Table of contents MASS 2100 FC300 FCT030 10 Operating Instructions 06 2017 A5E39789040 AA ...
Страница 36: ...Description 3 7 Theory of operation MASS 2100 FC300 FCT030 36 Operating Instructions 06 2017 A5E39789040 AA ...
Страница 90: ...Connecting 5 4 Device nameplates MASS 2100 FC300 FCT030 90 Operating Instructions 06 2017 A5E39789040 AA ...
Страница 258: ...Dimension drawings 12 4 Mounting bracket MASS 2100 FC300 FCT030 258 Operating Instructions 06 2017 A5E39789040 AA ...
Страница 478: ...Default settings B 15 Language MASS 2100 FC300 FCT030 478 Operating Instructions 06 2017 A5E39789040 AA ...
Страница 518: ...Index MASS 2100 FC300 FCT030 518 Operating Instructions 06 2017 A5E39789040 AA ...