Operation
72
Instructions for Use Carina SW 3.n
SPN-CPAP
Spontaneous Continuous Positive Airway Pressure
Suitable for patients with sufficient spontaneous
breathing. Spontaneous breathing at an increased
pressure level to raise the functional residual
capacity.
Set CPAP with the buttons for the ventilation
parameters:
– Positive end expiratory pressure »
CPAP
«
– O
2
concentration »
FiO
2
«
SPN-CPAP can be expanded with the following
ventilation parameters:
– Apnoea ventilation, see page 73.
With apnoea Ventilation switched on:
z
Set pressure limitation »
P
max
«.
z
Non-invasive ventilation with a hose system
with leak valve – using the »
Mask
« application
method, see page 75.
z
Setting alarm limits, see page 84.
Setting ranges
Ventilation parameter
Setting range
CPAP [mbar]
3 to 20
Inspiratory O
2
concentration FiO
2
[%]
21 to 100
Apnoea ventilation Apn. Vent.
On/Off
Pressure limitation P
max*
[mbar]
*
Only with apnoea Ventilation or volume guarantee switched on
LeakV: 5 to 40
ExpV: 5 to 50
529
530
Содержание Drager medical Carina
Страница 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 13: ...Instructions for Use Carina SW 3 n 13 Intended Use Intended Use Intended Use 14 Sub acute care definition 14 ...
Страница 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...