Operation
70
Instructions for Use Carina SW 3.n
SPN-PS
Spontaneous Pressure Support
For patients with sufficient spontaneous breathing.
Spontaneous breathing at an increased pressure
level to raise the functional residual capacity. Spon-
taneous breathing can be supported with Pressure
Support.
Set SPN-PS with the buttons for the ventilation
parameters:
– Pressure support »
Δ
P
supp
«
– Positive end expiratory pressure »
PEEP
«
– O
2
concentration »
FiO
2
«
– Sensitivity of »
Trigger
« (for synchronising
spontaneous breathing)
– Maximum inspiration time »
T
i max
«
– Pressure rise time »
Ramp
«
SPN-PS can be expanded with the following venti-
lation parameters:
– Apnoea ventilation, see page 73.
– Volume guarantee, see page 74.
With apnoea Ventilation or volume guarantee
switched on:
z
Set pressure limitation »
P
max
«.
z
Non-invasive ventilation with a hose system
with leak valve – using the »
Mask
« application
method, see page 75.
z
Setting alarm limits, see page 84.
526
527
528
Содержание Drager medical Carina
Страница 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 13: ...Instructions for Use Carina SW 3 n 13 Intended Use Intended Use Intended Use 14 Sub acute care definition 14 ...
Страница 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...