Instructions for Use Carina SW 3.n
123
Technical Data
Recommended separation distances
Recommended separation distances between port-
able and mobile RF-telecommunication devices
and the Carina.
Information regarding separation distances
(IEC 60601-1-2: 2001, table 205 and 206)
Max. P
EIRP
(W)
3 V/m distance
1
(m)
1
3 V/m distances to transmitters with frequencies from 150 kHz to 2.5 GHz, otherwise 1 V/m distance.
1 V/m distance
1
(m) Examples
0.001
0.06
0.17
0.003
0.10
0.30
0.010
0.18
0.55
0.030
0.32
0.95
e.g. WLAN 5250/ 5775 (Europe)
0.100
0.58
1.73
e.g. WLAN 2440 (Europe), Bluetooth
0.200
0.82
2.46
e.g. 5250 (not in Europe)
0.250
0.91
2.75
e.g. DECT devices
1.000
1.83
5.48
e.g. GSM 1800-/ GSM 1900-/ UMTS-
mobiles, WLAN 5600 (not in Europe)
2.000
2.60
7.78
e.g. GSM 900 mobiles
3.000
3.16
9.49
10.000
5.80
17.30
100.000
18.30
54.80
Содержание Drager medical Carina
Страница 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 13: ...Instructions for Use Carina SW 3 n 13 Intended Use Intended Use Intended Use 14 Sub acute care definition 14 ...
Страница 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Страница 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...