16. Load the paper into the paper slot on the cover.
17. Close the cover.
Figure 43 Closing the Cover
Replacing the Fuse
The fuse holder is located in the back panel between the power cord and the
power switch. It contains 2 fuses; both fuses are required. There are 2 grooves
above the fuse cover and 2 grooves below the fuse cover. Place the
screwdriver blade in the smaller groove to remove the cover.
Materials Required:
•
Flathead Screwdriver
•
Fuse: 250 V, T-1.25 A
1. Locate the 2 grooves on the top of the fuse holder.
2. Insert the blade of a small flat-head screwdriver into the smallest groove.
CAUTION
Set the power switch to Off. Disconnect the power cord from the wall
outlet. Disconnect the power cord from the system.
CAUTION
Ensure you place the screwdriver blade in the smaller groove. You
can damage the fuse block by placing the blade in the larger groove.