62
CLINITEK Analyzer Operator’s Guide
Mainten
ance
7. Rinse the test table and the table insert thoroughly with water.
8. Dry the test table and the table insert thoroughly with a soft cloth,
except for the white calibration bar.
9. Insert the test table and the table insert in the analyzer, as
described in
Weekly Cleaning of the Test Table and Test Table
Cleaning the Outside of the Analyzer
Always keep the outside of the CLINITEK analyzer clean and
free of dust.
To clean the outside of the analyzer, perform the following steps:
1. Power off the analyzer by pressing the on/off button for 2 seconds.
2. Wipe the outside (including the display) with a damp (not wet)
cloth and a mild detergent.
CAUTION
Rinse away all the solution residue, as any remaining solution
might affect the reagent pad chemistries.
BIOHAZARD
Wear personal protective equipment. Use universal
precautions. For recommended precautions when you work
with biohazardous materials, see
CAUTION
Do not use any type of solvent, oil, grease, silicone spray, or
lubrication on the analyzer.
Do not spray glass cleaner directly onto the screen.
Do not use laboratory wipes, such as Kimwipes, because they
might scratch the screen.
Prevent liquid from entering inside the printer compartment.
You could damage the analyzer or the printer.
Содержание CLINITEK Status+
Страница 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Страница 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Страница 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Страница 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Страница 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...