CLINITEK Analyzer Operator’s Guide
Introduction
Inserting the Batteries (optional)
To power the Clinitek Analyzer by batteries (optional), perform
the following steps:
1. Place the analyzer on its side.
2. Remove the battery cover on the bottom of the analyzer by
pressing down on the tab and pulling out the cover.
3. Place 6 new alkaline AA-size batteries into the battery
compartment.
4. Place the battery cover back on the compartment and turn the
analyzer back on its base.
Inserting the Test Table and Test Table Insert
To insert the test table and test table insert, perform the following
steps:
1. Insert the test table into the analyzer by holding it by the end
opposite the white calibration bar and with the white bar facing
up.
2. Push the test table into the analyzer, pushing it in just over
halfway.
3. Place the test table insert into the test table (see
).
CAUTION
Do not use batteries in the analyzer, if you attach the analyzer
to a CLINITEK Status connector. Leaving the batteries in the
battery compartment may corrode the batteries.
CAUTION
Do not push the test table fully into the analyzer. The test
table may become jammed and prevent the use of the
analyzer.
Do not touch the white calibration bar on the test table.
Damage to the calibration bar could affect the test results.
Содержание CLINITEK Status+
Страница 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Страница 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Страница 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Страница 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Страница 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...