CLINITEK Analyzer Operator’s Guide
111
System C
onf
iguration
4. Select the connectivity options.
5. Select
Next
.
The
Connectivity 2 of 2
screen displays.
6. Select
Serial Connection
.
7. Select
Edit serial settings
, and select the following options:
a. For
Baud
rate, select
9600
,
19200
,
57600
, or
115200
.
b. Select
Next
.
c. For
Parity
, select
None
,
Odd
, or
Even
.
d. Select
Next
.
e. For
Stop Bits
, select
1
or
2
.
8. Select
Done
4 times to return to the
Select Ready
screen.
Connectivity Option
Description
Screen 1 of 2
Connectivity Platform
•
Enabled
allows
communication between the
analyzer and the connector.
•
Disabled
(default) prevents
communication between the
analyzer and the connector.
Allow results to be sent
to PC
•
Enabled
(default) allows the
analyzer to send the test
results to a PC.
•
Disabled
prevents the
analyzer from sending the
test results to a PC.
Note
If you do not use a
CLINITEK Status connector,
Siemens recommends that you
select
Disabled
.