System Overview
Protective Measures
20
/ 44
AXD3-340.620.13.01.02
Operator Manual
On activation of one of the safety devices (shutdown devices) all system move-
ments stop and are blocked.
This means that one cannot move out from the blockage oneself.
If it is not possible to remove the obstruction causing the problem, call in the Sie-
mens Uptime Services.
Movements are possible again only after one of the safety devices has be deac-
tivated.
Possible collisions of the system with a
ceiling-mounted support
1
❏
System movements are possible only if the ceiling-mounted support is in its
park position or if the bypass switch is activated if necessary.
❏
Prior to actuating system movements with the bypass switch activated, verify
that no collision can take place with the ceiling-mounted support, the X-ray
tube assembly or the collimator. (No collision monitoring.)
Measures for avoiding equipment damage
◆
Before activating system movements, especially tilting the table, make sure
that the movement range is free of obstructions.
◆
Move especially monitor support systems, operating consoles, gurneys, beds
and instrument tables out from the tilting range of the table and remove
chairs, steps, stands, waste containers and similar objects from the move-
ment area. (No collision monitoring.)
◆
Do not place any objects or consumable material on the cover of the table sup-
port, on the spotfilm device and on the longitudinal guides of the stand car-
riage.
– Considerable forces which can damage these objects in the area of move-
ment of the systems arise during movements of the spotfilm device or of
the tabletop.
◆
Do not place any loose objects anywhere on the table.
– These objects could fall down when the table is tilted, causing injury or
damage.
1
optional
Warning
If system movements are possible even though the ceiling-mounted support is
not in the park position and the bypass switch is not activated,
immediately press the emergency shutdown button and notify the
SIEMENS Uptime Service.
Содержание AXIOM Iconos R200
Страница 2: ......
Страница 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Страница 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Страница 90: ...Examination Transferring and Positioning the Patient 6 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Страница 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Страница 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Страница 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...