
34
ADVIA 1800 Operator’s Guide
Reaction tank
The reaction tank contains non-reactive oil, which keeps the temperature of the liquid in
the reaction tray (RRV) cuvettes at a constant 37
°
C
±
0.5
°
C. A heater and a thermostat
control the temperature.
1
Reaction Tank
2
Two Liquid Surface Level
Sensors
Figure 1-25. Reaction tank
Spectrophotometer
The spectrophotometer measures the amount of light absorbed at 14 specific wavelengths
by liquids contained in reaction cuvettes.
Every six seconds, the reaction tray (RRV) moves cuvettes containing reaction liquid
(sample and reagent) in front of a halogen lamp, which sends light through the cuvettes.
Each time, a different wavelength is measured.
The photometer then measures the absorbance based on the lamp energy and the optical
density of the cuvettes. This process is repeated for as many times and wavelengths as
required by the assay conditions.
A cooling tank maintains the lamp temperature.
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...