219
Appendix F – Flags
Alarm (result) flags
Flags
Flag name
Reason
RealTime
Monitor
Review/
Edit
Patient
Report
Sent to
LIS
*
Variance
Photometric tests Assay Data
imprecise
ISE tests—calibration value
error
Y
N
Y
Y
N
N
/////
Overflow
Calculation error. Do not
report as a result. Do not
interpret as a result. This
mark displays in the result
field.
cal error
cal error cal error
A
Clot error
A clot is detected
N
N
N
b
Pack switch
A pack was switched during
calibration
N
Y
C
Calibration
error
Photometric tests—no or
faulty calibration
ISE tests—ISE calibration
range error
Y
cal error
Y
cal error
N
cal error
d
Photometric tests-Abnormal
reaction absorbance (lower)
ISE tests—dilution error,
refernce electrode value is
less than 500
Y
N
Y
Y
N
N
D
Absorbance
limit (lower)
Lamp energy low during cell
blank measurement check
Photometric tests—Abnormal
reaction absorbance (lower)
Y
Y
N
E
Edit
A result was edited by the
operator
Y
N
__
F
Temperator
The reaction bath
Y
Y
N
Содержание ADVIA 1800
Страница 1: ...11221680 Rev A 2015 12 Operator s Guide ADVIA 1800 Chemistry System ADVIA 1800 Chemistr y System ...
Страница 102: ...102 ADVIA 1800 Operator s Guide ...
Страница 108: ...108 ADVIA 1800 Operator s Guide ...
Страница 182: ...182 ADVIA 1800 Operator s Guide ...
Страница 188: ...188 ADVIA 1800 Operator s Guide ...
Страница 194: ...194 ADVIA 1800 Operator s Guide ...
Страница 198: ...198 ADVIA 1800 Operator s Guide ...
Страница 204: ...204 ADVIA 1800 Operator s Guide ...
Страница 212: ...212 ADVIA 1800 Operator s Guide ...