Parameter assignment
7.22 Measurement verification (P709 to P713)
SITRANS Probe LR with mA/HART
60
Operating Instructions, 05/2019, A5E32337711-AC
Enter a value slightly greater than the vessel’s maximum emptying rate, in Units (P005) or
percent of Span (P007) per minute.
P003 value
Meters/minute
1
0.1
2
1
3
10
7.22
Measurement verification (P709 to P713)
Note
Factory settings are indicated by an asterisk (*) in the parameter tables, unless explicitly
described.
7.22.1
P709 Damping filter
Stabilizes the reported level within the Echo Lock Window (P713) in the event of level
fluctuations (for example, a rippling or splashing liquid surface). The value is in seconds, and
depends on the number of seconds it takes the device to reach 63% of a step value change
in reading.
Values
Range
0 to 100 seconds (0=off)
Default
10.0 seconds
Altered by
P003 Measurement response
Related parameters
P007 Span
P713 Echo lock window
The value is automatically altered when Measurement Response Speed (P003) changes.
The higher the value entered, the greater the range of stabilized fluctuation.
7.22.2
P711 Echo lock
Note
Ensure the agitator is always running while SITRANS Probe LR is monitoring the vessel, to
avoid stationary blade detection.