Example:
A flow velocity of 3 m/s (10 ft./sec.) (V) in a sensor with a diameter reduction from DN 100 to DN
80 (4" to 3") (d
1
/d
2
= 0.8) gives a pressure drop of 2.9 mbar (0.04 psi).
4.1.3
Mounting the sensor
1. Install gaskets.
2. Ensure connection flange has a smooth surface and is in line with sensor.
Gaskets are recommended but not included in water meter delivery.
Installing/Mounting
4.1 Sensor installation
SITRANS FM MAG 8000
22
Operating Instructions, 03/2020, A5E03071515-AH