Functions
7.1 Selecting Flow Units
FUP1010 IP67 Portable
80
Operating Instructions, 03/2012, A5E02951522-02
8.
At the Terminal screen, press <ENTER> and the Data Display mode appears.
9.
If not, to enter the Data Display mode type MENU and then press <Ctrl L>.
Table 7- 2
Totalizer Controls (the "n" in <Fn> = channel number)*
Key
PC #
Command
Description
F1
F2
F3
F4
CLRTOT
(also clears
overflow)
Resetting the Totalizer registers clears all total data accumulated during operation.
Note: In Dual Path mode, the Totalizer operates only on the virtual system channel
(Ch 3). Therefore in this case, the CLRTOT trigger would be <F3> <1>.
Commands that can be invoked from Si-Ware or HyperTerminal:
Terminal Command: CLRTOT 1.
F1
F2
F3
F4
NOTOT
(Totalizer Freeze)
Invoking the NOTOT command disables the Totalizer. Totalization will not resume
until you repeat the <Fn> <2> key sequence. When you activate NOTOT, an N
precedes the TOTAL symbol (i.e. [NTOTAL]) on the LCD Screen.
Commands that can be invoked from Si-Ware or HyperTerminal:
Terminal Command: NOTOT 1 = Stop Totalizer
Terminal Command: NOTOT 1 = Start Totalizer
F1
F2
F3
F4
LAPTOT
(Totalizer
snapshot)
The LAPTOT command freezes the Totalizer screen display. However, the flow
meter will continue to update its internal registers. The flow meter will show the
current total when you repeat the <F1>< 3> key sequence. When you activate
LAPTOT, an L precedes the TOTAL symbol (i.e., [LTOTAL]) on the HyperTerminal
screen.
F1
F2
F3
F4
CLEAR
(Batch/Tot
register)
Clears the Batch/Sample Totalizer register. The flow meter maintains a separate
Totalizer register for Batching or Sampling applications but cannot be accessed
directly. It is used for relay control only. If you assign the system relay to this
function, a momentary (200 mS) relay pulse occurs whenever the BATCHTOT
register accumulates a specified liquid quantity. In the [Batch/Sample Tot] menu cell
the required total flow volume is entered to activate the relay,. This numeric entry
must reflect the selected flow total units. The [Totalizer Scale] menu cell shows the
applicable flow total units. The sign of the Batch/Sample Total determines positive or
negative accumulation.
F1
F2
F3
F4
CLEAR
(Makeup Latch)
Clears the Makeup Latch. Refer to the Span Data menu [Set Alarm Levels] and then
the [Makeup Latch] On / Off option.
*Use the <F1> key as the "Lead-in command" for 4-Path Totalizer operations.
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......