Description
3.3 Theory of Operation
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
15
Typical Industries Serviced
●
HVAC (Hotels, Airports, Government)
●
Power Generation (Nuclear, Fossil, and Hydro)
●
Chemical Processing
●
Food and Pharmaceutical
●
Aircraft Avionics and Ground Support
●
Water and Wastewater
●
Aerospace
●
Automobile Manufacturing
●
Hydrocarbon Industries
3.3
Theory of Operation
The liquid flow meter relies on the MultiPulse transit. Two WideBeam ultrasonic sensors per
measuring path, alternating as transmitter and receiver, are used to interrogate the liquid
flowing within the metering section. The resulting time of arrival for each direction of transmit
(upstream and downstream) is then measured using a highly accurate and stable digital
signal processing method.
Using this detection scheme, the flow meter is capable of resolving the relative transit-time
difference (dT) to within ±100 psec. Considering typical liquid flow transit-time differences
ranging from 1x10
4
to 1x10
6
psec, the flow meter is capable of providing an exceptional turn-
down ratio. The flow meter also incorporates a correlation technique which enables the
system to detect very high flow velocities with the same high degree of resolution. The
ultrasonic sensors are designed with sufficient beam divergence characteristics to insure that
the receive sensor will always have sufficient signal to maintain operation under conditions of
high beam blowing associated with high flow velocities.
With accurate signal arrival time available, the flow meter can compute the raw flow velocity
from the measured upstream and downstream transit times.
Содержание 7ME3510
Страница 2: ......
Страница 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 68: ...Connecting 5 3 Zero Flow Adjust Menu FUP1010 IP67 Portable 66 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 116: ...Functions 7 11 Reflexor FUP1010 IP67 Portable 114 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 144: ...Technical data 11 1 Technical Data FUP1010 IP67 Portable 142 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 159: ...1010WP 7 21614 C ...
Страница 160: ...1010WP 7 21614 C ...
Страница 161: ...1010WDP 7 21614 C ...
Страница 162: ...1010WDP 7 21614 C ...
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Страница 168: ......
Страница 169: ...21614 C ...
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ...21614 C ...
Страница 175: ...21614 C ...
Страница 176: ...21614 C ...
Страница 177: ...21614 C ...
Страница 178: ...21614 C ...
Страница 179: ...21614 C ...
Страница 180: ...21614 C ...
Страница 181: ......
Страница 182: ......
Страница 183: ...21614 C ...
Страница 184: ...21614 C ...
Страница 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Страница 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Страница 187: ...1015BC 1 8 21614 C ...
Страница 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Страница 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Страница 196: ......
Страница 197: ......
Страница 198: ...L TEMPERATURE BELOW 225 F ONLY SUITABLE FOR AMBIENT TEMPERATURE GLAND SUPPLIED WITH CABLE IS 21614 C PART No L ...
Страница 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Страница 213: ......