Required forms
• Delivery note
• Return goods delivery note (
https://www.siemens.com/processinstrumentation/
with the following information:
– Product (item description)
– Number of returned devices/replacement parts
– Reason for returning the item(s)
• Decontamination declaration (
With this declaration you warrant "that the device/replacement part has been carefully
cleaned and is free of residues. The device/replacement part does not pose a hazard for
humans and the environment."
If the returned device/replacement part has come into contact with poisonous, corrosive,
flammable or water-contaminating substances, you must thoroughly clean and
decontaminate the device/replacement part before returning it in order to ensure that all
hollow areas are free from hazardous substances. Check the item after it has been cleaned.
Any devices/replacement parts returned without a decontamination declaration will be
cleaned at your expense before further processing.
8.3
Keying in sensor data
The following step-by-step procedure only applies to:
• FUS060 transmitters with SONOKIT sensors
• Replacement transmitters
NOTICE
SONO 3100 and SONO 3300
All FUS060 transmitters calibrated with sensors SONO 3100 or SONO 3300 are delivered with
sensor data preset from factory. The data should not be changed
The step-by-step procedure describes the required parameters/data to be entered and the
sequence of entering them in order to configure the transmitter correctly for a given pipe
application.
Note
The pipe data are taken from the SONOKIT sensor geometry measurement report or from
another sensor data sheet or calibration report.
Service and maintenance
8.3 Keying in sensor data
SITRANS FUS060 with PROFIBUS PA
72
Operating Instructions, 09/2021, A5E02225054-AD