Service programs
5 - 5
Siemens-Elema AB
SPR8-215.898.02
Page 5 of 8
MOBILETT Plus/ Plus E
Solna, Sweden
Rev. 01
01.02
SPS-UD
The kV is changed in steps by briefly depressing the "kV + or kV -" buttons. By keeping
the buttons pressed, the kV value will be continuously stepped up or down.
The generator filament values are self-adjusted to produce a kV value with minimum devi-
ation. If the software was optimized, a filament value for a specific kV value was indicated
with the letter "A" in the extreme right position on the mAs display.
P08 - Display the message history
5
The history contains two parts. The first is the appearance number and the second the
actual message. The appearance number will be displayed in the "kV" field. The corre-
sponding message will be displayed in the "mAs" field. Pressing the "kV +" button will dis-
play the next higher number and corresponding message and vice verse with "kV -". The
program starts by showing the latest message in the history. If there are no messages the
display will show:
The latest message has the highest number, for example:
To exit P08, use the "lamp" button.
P09 - Clear the message history
5
Activate the program with the "lamp" button. The display shows:
To clear the message history, press the "mAs +" button for 5 seconds. The program con-
firms that the history is clear:
After the message clears, the program is ended.
kV mAs
0
Error = Message history empty
kV mAs
20 CAL13 = There are total 20 new messages, number 20 is "CAL 13".
When this program is run, the history will be cleared and the file
will be empty. Use program P08 and record all messages before
using program P09.
kV
mAs
P09 CLEAr
kV
mAs
0
Error = Message history empty
NOTE