Xprecia Stride
User Guide
5. Performing troubleshooting and maintenance
Learning about system messages
Stride stores and lets you retrieve a minimum of 300
system messages, comprised of warning messages and
error messages:
Warning message
Stride detects a problem, displays
a message, and once you correct the problem, you can
continue the test. These messages are stored in the
Events log.
Error message
Stride detects a problem, displays a
message, but you can’t continue testing. These messages
are stored with the test results.
Viewing the system messages
INSTRUCTIONS
1. On the Home screen, tap
.
2. Tap
.
Operator 3245W
9:58am 10.05.2015
02-07
01-02
01-06
01-05
9:46am 10.05.2015
Operator 2980W
9:42am 10.05.2015
9:36am 10.05.2015
10:00 am
10.05.2015
The most recent system messages display at the top of the list,
showing the time, date, warning, or error code.
Troubleshooting system messages
Warning messages
CODE MESSAGE
PROBABLE CAUSE
POSSIBLE SOLUTION
01-01
Barcode scan error
Stride encountered an unknown error
while scanning.
Try to scan again. If the error continues, turn Stride off
and on.
01-02
Battery Low
Your batteries are running out.
Replace with 3 new AA batteries.
01-04
Invalid Login
An incorrect Operator ID (OID) was
entered.
Try reentering the Operator ID (OID).
01-05
Invalid OID
An incorrect Operator ID (OID)
was entered.
Try reentering the Operator ID (OID).
01-06
Invalid PID
An incorrect Patient ID (PID)
was entered.
Try reentering the Patient ID (PID).
01-07
LQC Expiration date exceeded
Your lot of control solution has
reached its expiration date.
Verify the Stride date setting is correct and then redo
the test using a different control lot.
01-08
OID delete failed
Stride encountered an error trying to
delete the Operator ID (OID).
Make sure you have administrator rights to delete an
OID and that the OID is valid.
01-09
OID list full
The operator list is full.
Remove one or more operators from the list.
01-10
Test strip port protective cap missing
The test strip port protective cap was
not installed properly.
Align tab in cap with slot and slide cap until it
clicks into place.
01-11
Test strip inserted early
The test strip was inserted before Stride
requested.
Remove test strip and insert when Stride prompts you.
01-12
Test strip and vial do not match
You may have scanned a test strip from
a different lot.
Verify the test strip is from the correct vial and rescan.
If the error continues, cancel the test and repeat using
a new test strip from the vial.
Содержание XPRECIA STRIDE
Страница 2: ...XPRECIA STRIDE USER GUIDE ...
Страница 4: ...1 Start here ...
Страница 6: ...2 1 Exploring Stride ...
Страница 20: ...30 2 Per forming a patient test ...
Страница 26: ...42 3 Per forming a LQC test ...
Страница 30: ...50 4 Ch anging the settings ...
Страница 35: ...60 5 Performing troubleshooting and maintenance ...
Страница 43: ...76 Appe ndix A Specifications ...
Страница 45: ...80 Appendix B Safety ...
Страница 48: ...86 A ppendix C Operational theory ...
Страница 51: ...92 Appendix D Symbols ...
Страница 53: ...96 Appendix E Glossary ...