2. Performing a patient test
Fingerstick sample collection method
Use the following instructions to obtain a fingerstick
patient blood sample.
Requirement
Only use auto-disabling, single-use
lancing devices with Stride.
Collecting a fingerstick blood sample
INSTRUCTIONS
1. To stick the finger, firmly place the lancet against the
finger and press the lancet trigger.
2. Gently squeeze from the base of the finger to form a
round drop of blood. If the blood smears or runs, wipe
it off with a tissue and gently squeeze another round
drop of blood.
Requirement
The drop of blood should be about the
same size as the test strip target area (a minimum
of 6 μL in volume). Low sample volume will cause
an error message.
3. Return to page 36 to follow steps 12–16.
IMPORTANT
Apply the drop of blood to the test strip target area within
15 seconds of lancing the fingertip.
Viewing patient test results
After you perform a patient test, the results display.
INR
10:00 am
PT
10.05.2015
1.2
@
9:58 am
10.05.2015
Outpatient 12K9
Operator 2980W
The following information appears on the screen: INR value,
time and date, Patient ID (PID), and Operator ID (OID) (if entered).
Understanding results
Results display in the International Normalized Ratio (INR).
Because desired ratios may vary depending upon the
clinical practice and test methodologies, the optimum
therapeutic range for this method should be established by
each user.
Each lot of Xprecia System PT/INR test strips is calibrated
to a reference lot of human recombinant thromboplastin
traceable to the World Health Organization International
Reference Preparation.
1
INR =
(
[Patient Prothrombin Time (sec)]
)
ISI
[Mean Normal Prothrombin Time (sec)]
The calculation is performed with an ISI of 1.0 and a typical
Mean Normal Prothrombin Time of 12.0 seconds.
1. WHO Expert Committee on Biological Standardization. Forty-eighth Report. Geneva,
World Health Organization, 1999 (WHO Technical Report Series, No. 889)
Содержание XPRECIA STRIDE
Страница 2: ...XPRECIA STRIDE USER GUIDE ...
Страница 4: ...1 Start here ...
Страница 6: ...2 1 Exploring Stride ...
Страница 20: ...30 2 Per forming a patient test ...
Страница 26: ...42 3 Per forming a LQC test ...
Страница 30: ...50 4 Ch anging the settings ...
Страница 35: ...60 5 Performing troubleshooting and maintenance ...
Страница 43: ...76 Appe ndix A Specifications ...
Страница 45: ...80 Appendix B Safety ...
Страница 48: ...86 A ppendix C Operational theory ...
Страница 51: ...92 Appendix D Symbols ...
Страница 53: ...96 Appendix E Glossary ...