Xprecia Stride
User Guide
2. Performing a patient test
Warnings and cautions for running a patient test
Always follow these safety and usage guidelines for the most accurate results:
‒ Don’t take test strips internally or drink control solutions.
‒ Always store the test strips in the original test strip vial with the cap closed.
‒ Use the test strip within 5 minutes of when you remove it from the test strip vial.
‒ Use only fresh capillary (fingerstick) whole blood for tests.
‒ Hematocrit range of 22–52% doesn't significantly affect test results.
‒ Apply the blood sample to the test strip within 120 seconds after the Apply Sample screen displays.
‒ Always wear protective gloves and follow your facility’s policies and procedures when performing tests involving
biological samples and control solutions.
‒ Only apply blood to a test strip after it's inserted in, and blood is requested by, Stride.
‒ Never use bent, scratched, or damaged test strips.
‒ Don’t scan a test strip barcode and then use a different test strip from another vial.
‒ Each test strip is single-use only; never perform a second test using the same test strip.
‒ Don’t touch or move the test strip after you apply the drop of blood. Don't move the test strip during the test.
‒ Never add more blood to the test strip after the test has begun.
‒ Refer to the
Xprecia System PT/INR Strips Instructions For Use
for more information on test strips.
‒ A best practice is to make sure there is adequate lighting to conduct a test.
‒ Don't conduct tests when connected to Xprecia DMS.
‒ Always thoroughly wash and dry your hands and put on a new pair of gloves for each patient test.
‒ See
Biosafety in Microbiological and Biomedical Laboratories (BMBL)
found at
http://www.cdc.gov/biosafety/publications for more safety information.
Disposing of waste
Observe the following guidelines when disposing
biohazardous waste:
‒ Dispose of used lancets in an approved sharps container.
‒ Dispose of used test strips in an approved
biohazard container.
‒ Always follow your facility’s biohazard disposal policies.
When ejecting a used test strip, always
point Stride down facing
your biohazard container before you press the Test Strip
Eject
button.
10:00 am
10.05.2015
Содержание XPRECIA STRIDE
Страница 2: ...XPRECIA STRIDE USER GUIDE ...
Страница 4: ...1 Start here ...
Страница 6: ...2 1 Exploring Stride ...
Страница 20: ...30 2 Per forming a patient test ...
Страница 26: ...42 3 Per forming a LQC test ...
Страница 30: ...50 4 Ch anging the settings ...
Страница 35: ...60 5 Performing troubleshooting and maintenance ...
Страница 43: ...76 Appe ndix A Specifications ...
Страница 45: ...80 Appendix B Safety ...
Страница 48: ...86 A ppendix C Operational theory ...
Страница 51: ...92 Appendix D Symbols ...
Страница 53: ...96 Appendix E Glossary ...