57
8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
OPERATION
4
4.15
Display all changeable parameter values (command PARAM)
Displays the current values of all parameters that can be set by the user.
Syntax:
PARAM
<cr>
Example:
4.15.1
Measure signal level (command SIL)
Tests the signal level. The laser signal intensity is compared to the signal intensity (factory
calibration) originally set. The result is shown as 0 ... 100% of the original signal intensity
set. This allows measuring the contamination on optical surfaces.
Syntax:
SIL
<cr>
Example:
4.15.2
Display statistic information (command STATS)
Displays statistic information.
Syntax:
STATS
<cr>
Example:
>param
Customer Interface
SERI : 19200 8 NONE 1
ECHO : ON
SMODE : STOP
Service Interface
SERI : 115200 8 NONE 1
ECHO : ON
SMODE : STOP
Common Serial parameters
ADDR : 0
INTV : 1 S
FORM : F0
Analog Output
OUTMINO2 (%) : 0.000
OUTMAXO2 (%) : 25.000
I4 : 1
NONFATALI (mA) : 3.000
FATALI (mA) : 3.000
Relay Output
RELAY_MODE : FAULT_ALARM
LO POINT (%02) : 10.0
HI POINT (%02) : 11.0
Measurement parameters-
INSTALLATION : Process measurement
PRESSURE(bar) : 1.000
H2O (g/m3) : 50
CO2 (vol%) : 20
>sil
Signal level is 100% compared to signal level at factory
>stats
All cleared : 2006-01-18 13:40:04
Uptime (h) : 140
Resets : 7
O2 max:21.06
O2 min : 4.91
Tg max : 29.71
Tg min : 23.39
Ti max : 32.53
Ti min : 24.55