8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
SPECIFICATIONS
10
10.4.8
Connection lines
Connection line from TSA151 to TRANSIC151LP
Connection line from AO and DO on TSA151 to the isolation amplifiers
Connection line from 24 V PELV power supply unit to TSA151
Maximum line length
5 m
Maximum line cross-section
0.5 mm
2
(flexible wire with ferrule)
Cable specifications
Design
3 x 2 x 0.5 mm
2
, shielded (shielding braid) and twisted as pair
Inductance Lc
≤
1 mH/km
Capacity Cc
≤
120 nF/km
Test voltage min.
≥
500V AC / 750V DC
Conductor diameter
≥
0.1 mm
[1]
Outer diameter
7 … 12 mm (M20 cable gland in scope of delivery)
Temperature range (flexible)
−
20 °C … +80 °C
Wire identification
In accordance with DIN47100
[1] Also valid for single wires of a fine-stranded conductor.
▸
Identify intrinsically safe power circuit (e.g., with light blue cable sheath).
For maximum load of 200 Ohm
Max. line length
400 m with 200 Ohm load (AO)
Max. line cross-section
0.5 mm
2
(flexible wire with ferrule)
For maximum load of 150 Ohm
Max. line length
800 m with 150 Ohm load (AO)
Max. line cross-section
0.5 mm
2
(flexible wire with ferrule)
Cable specifications
Design
2 x 2 x 0.5 mm
2
, shielded (shielding braid) and twisted as pair
LC
≤
1 mH/km
Cc
≤
120 nF/km
Test voltage min.
≥
500V AC / 750V DC
Conductor diameter
≥
0.1 mm
[1]
Outer diameter
7 … 12 mm (M20 cable gland in scope of delivery)
Temperature range
−
20 °C … 80 °C
[1] Also valid for single wires of a fine-stranded conductor.
– With line cross-section = 1 mm
2
Max. 100 m
– With line cross-section = 1.5 mm
2
Max. 150 m
Min. input voltage required on TSA151 input as well as losses on the line are considered.
Cable specifications
Design
2 x 1.0 mm
2
oder 2 x 1.5 mm
2
Temperature range
−
20 °C … +80 °C
Outer diameter
5 … 9 mm (M20 cable gland in scope of delivery)
A 24V PELV power supply unit with adjustable output voltage allows compensating
losses on the line and therefore larger distances. Note that this is the responsibility of
the operator/installer.