75
8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
SPECIFICATIONS
10
10.4.4
TRANSIC151LP electrical connection
(Terminals) TRANSIC151LP connection data
10.4.5
TRANSIC151LP safety-related technical characteristic data (IECEx/ATEX)
TRANSIC151LP voltage supply input
TRANSIC151LP analog output
TRANSIC151LP switching output (NAMUR)
Supply voltage range
7.5 V (supply only via TSA151)
Typical current consumption
360 mA
Typical power input
2.7 W
Possible line cross-section on the
connection terminals (zone Ex-ib)
Rigid:
●
Min. 0.5 mm
2
●
Max. 1.5 mm
2
Flexible:
●
Min. 0.5 mm
2
●
Max. 1.5 mm
2
Flexible with ferrule with plastic sleeve:
●
Min. 0.5 mm
2
●
Max. 0.5 mm
2
AWG:
●
Min. 20
●
Max. 16
Connection type: Spring force
Insulation stripping: 9 mm
Gas group
IIB
Maximum voltage Ui
8.25 V
Maximum current Ii
650 mA
Maximum performance Pi
5.37 W
Maximum inductance Li
≈
3.5
μ
H
Maximum capacity Ci
≈
50
μ
F
Maximum voltage Uo
8.25 V
Maximum current Io
121 mA
Maximum performance Po
250 mW
Maximum inductance Lo
5 mH
Maximum capacity Co
1 uF
Maximum voltage Ui
10 V
Maximum current Ii
10 mA
Maximum performance Pi
100 mW
Maximum inductance Li
≈
3.5
μ
H
Maximum capacity Ci
≈
132 nF