67
8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
SPECIFICATIONS
10
10
Specifications
10.1
Compliances
The technical design of this device complies with the following EU
directives and EN standards:
●
Directive 2014/30/EU (Low Voltage Directive, Harmonization of the
laws)
●
Directive 2011/65/EU (Restriction of the use of certain hazardous
substances in electrical and electronic equipment – “RoHS”)
●
EN 61326-1 (Electrical equipment for measurement, control and laboratory use – EMC
requirements – Part 1: General requirements (IEC 61326-1:2012))
●
EN 50581 (Technical documentation for the assessment of electrical and electronic
products with respect to the restriction of hazardous substances)
Electrical protection
●
Protection class III: Low voltage protection
●
Electrical safety in accordance with DIN EN IEC 61010-1:2011.
●
Supply with low voltage protection PELV (in accordance with EN 60204).
A 24V PELV power supply unit (with Um = 60V) is mandatory for supplying the TSA151 to
ensure electrical safety.
10.2
Ex Approvals
The TRANSIC151LP is suitable for the following zones
In accordance with ATEX (EN60079-10) and
In accordance with IECEx (IEC60079-10)
●
Gas:
– Measuring probe: Category 1G EPL Ga (Zone 0)
– Transmitter / voltage supply unit: Category
2G EPL Gb (Zone 1)
●
Dust:
– Transmitter / voltage supply unit / measuring probe:
Category 2D EPL Db (Zone 21)
The TRANSIC151LP may be used in zones in which inflammable or explosive gases in
groups I, IIA and IIB or inflammable dust in groups IIIA, IIIB and IIIC are present.
●
Temperature class:
– Gas: T4 (max. surface temperature 135°C)
– Dust: T85°C
Further information on Ex approval
see “Explosion protection according to ATEX and
.
WARNING:
Ex certification becomes void for unallowed power supply
▸
Only use the TSA151 voltage supply unit to run the TRANSIC151LP.
▸
Only use a 24V DC PELV power supply unit on the TSA151 voltage supply unit. (
“Special operating conditions”, page 10
▸
Observe the “X” marking on the TSA type plate,