7.3.8.5
Setting limits
The threshold is set using limits. In doing so, the digital signal is set to
ON
or
OFF
. Not
available for the edge height change switching mode.
Table 16: Upper limit of setting options
Device
Possible values
Default value
OD5000-C15xxx
±
1.00 mm
+1.0 mm
OD5000-C30xxx
±
5.00 mm
+1.0 mm
OD5000-C85xxx
±
20.00 mm
+1.0 mm
OD5000-C150xxx
±
40.00 mm
+1.0 mm
Table 17: Lower limit of setting options
Device
Possible values
Default value
OD5000-C15xxx
±
1.00 mm
-1.0 mm
OD5000-C30xxx
±
5.00 mm
-1.0 mm
OD5000-C85xxx
±
20.00 mm
-1.0 mm
OD5000-C150xxx
±
40.00 mm
-1.0 mm
7.3.8.6
Setting the hysteresis
If the measured value rises slightly above or drops slightly below the threshold, this can
lead to the input and output repeatedly switching on and off within a short space of
time. This can be prevented by assigning a defined tolerance to the measured value via
the hysteresis setting, beyond which the device switches.
NOTE
The limits and the hysteresis can also be set by moving the corresponding marker in the
graphic (blue: limits, yellow: hysteresis).
Default value is 0.1 mm.
Hysteresis
2
ON
OFF
Upper
limit
1
Time
3
Figure 54: Hysteresis
1
Upper threshold
2
Hysteresis
3
Time
7.3.8.7
Setting the switch-on delay (on delay)
This setting is used to delay the switching on of the output by a set amount of time if
the measured value exceeds the threshold.