72
8018478/YWL5/2018-06| SICK
O P E R A T I N G I N S T R U C T I O N S | NAV245
Subject to change without notice
10
TECHNICAL DATA
10
Technical data
10.1
NAV245 data sheet
Characteristic
NAV245
Landmark detection
[1]
Measuring range
0.5 to 50 m on reflectors
Usable scanning angle
270
for reflectors which are fully hit
Angular resolution
0.001
on reflectors
Systematic distance measurement error
Typically:
10 mm for a distance of up to 30 m
Statistical distance measurement
error (1 Sigma)
Typically: 6 mm
Systematic angle measurement error
Typically: ±0.04° for distances between 2 and 30 m
Temperature drift
Typically:
0.1 mm/K
Contour data
[2]
Measuring range
0.5 to 50 m
Sensing range
18 m at 10 % remission
50 m at 80 % remission
Usable scanning angle
270
Angular resolution (increment)
0.25
Scanning frequency
25 Hz ± 5 %
Measurement resolution distance
1 mm
Systematic error distance
Typically: ± 30 mm
Maximum: ± 50 mm
Statistical distance error (1 Sigma)
Typically: 12 mm
Temperature drift
Minimum:
0 mm/K
Maximum:
0.32 mm/K
General data
Beam divergence
15 mrad
Laser diode (wavelength)
Infrared light (
λ
= 905 nm)
Laser class of the device
[3]
Class 1 as per IEC 60825-1:2014, eye-safe, complies
with 21 CFR 1040.10 and 1040.11 expect for
deviations pursuant to Laser Notice 50, dated
June 24, 2007.
RS-232 data interface
9,600 to 115,200 Bd, default: 115,200 Bd
Data format
Variable, default: 8 data bits, 1 stop bit, even parity
Ethernet data interface
10/100 Mbit/s, TCP/IP
Digital switching outputs
Potential-free semiconductor output (see chapter 5.4.4)
I max = 100 mA at 11 to 30 V DC
Current limit: 100 to 200 mA
Operating voltage
10.8 V DC to 30 V DC
Protection class
III according to EN 61140 (2002-03)
Enclosure rating
IP 67 according to EN 60529 (1991-10); A1 (2002-02)
NAV245-10100S01: IP 65 according to EN 60529
(1991-10); A1 (2002-02)
Vibration check
According to IEC 60068-2-6 (1995-04)