Clearing Malfunctions
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
8.7
Error messages and possible causes
Notes on the error messages:
●
This Table also contains recommended solutions that can only be performed by spe-
cially trained personnel.
●
If fault persists: Contact SICK Customer Service.
Initiator Code Error text
Classifi-
cation
Description
Possible clearance
System S001 Communication error
Failure
Communication fault between system and sub-
assembly.
Check cable connections in
system cabinet.
Restart the device.
S002 Configuration error
Configuration error, found module does not cor-
respond to that of the nominal configuration
Please contact SICK Customer
Service
S003 Temperature
Temperature outside limit value.
Check temperature parame-
ters in menu 3.2
Diagnosis/
Temperatures
.
Check heating circuits.
S004 EEPROM
Internal fault.
Please contact SICK Customer
Service
S005 Analog input too high
Current at analog input above 22 mA
Check parameter settings of
connected analog output and
adapt, if required.
S006 Voltage range
The input range of an analog input was
exceeded.
S007 Check sum error
S008 Cell flow
Flow through cell is outside tolerance range
(150 - 400 l/h).
Check induction air.
Check pressure parameter P1/
P2 in SOPAS ET menu
Diagno-
sis/Sensor values/Pressures
.
Check sample gas outlet for
blockages.
S009 Cell pressure
Cell pressure is outside tolerance range.
S010 Ambient pressure
Ambient pressure is outside tolerance range.
Check pressure sensor.
S011 Ambient temperature
Temperature in cabinet is outside tolerance
range.
Check air conditioner.
S012 Vaporizer temperature
Vaporizer temperature is outside tolerance
range.
Check vaporizer temperature in
SOPAS ET menu
Diagnosis/
Sensor values/Temperatures
.
Check communication between
the test gas generator and the
device by means of logbook
error message ETH1. If this
error occurs, there is no com-
munication between the test
gas generator and the device.
Check cables.
S013 Lamp energy
Lamp energy is outside tolerance range.
Check lamp energy in SOPAS ET
menu
Diagnosis/Sensor val-
ues/Lamp
.
S014 Lamp ignition
Lamp ignition failure.
Reinitialize device in menu
1.2.4
Maintenance/Change
operational state
.