69
8013741/16LH/V1-7/2020-04| SICK
O P E R A T I N G I N S T R U C T I O N S | MCS300P Ex
Subject to change without notice
CLEARING MALFUNCTIONS
8
8.7
Error messages and possible causes
Source
Code
Error text
Classifica-
tion
Description
Possible clearance
[1]
[1] This Table also contains recommended solutions that should only be performed by specially trained personnel.
System
S001 Temperature T1 too high
Failure
If T1 > (nom.temp + param.limit)
Check the heater
S002 Temperature T2 too high
If T2 > (nom.temp + param.limit)
S003 Temp. 1 not reached
After 60 min. in current operation: 15 min
Refer also to
S004 Temp. 2 not reached
S005 Temperature sensor 1
OVO (HC3X) signals that the input range of an
analog input (temperature sensor) was exceeded
S006 Voltage range
OVO (HC3X) signals that the input range of an
analog input (temperature sensor) was exceeded
Please contact SICK Customer
Service
S007 Check sum error
OVO (HC3X) signals that the input range of an
analog input (temperature sensor) was exceeded
S008 Chopper signal missing
This message is first displayed when the receiver
has set the corresponding bit 5 times (seconds) in
succession
S009S
010
S011
Motor x: Ref.pos.incorrect
Filter wheel motor x does not detect the reference
position
S012 No emitter detected
No emitter detected
Check the voltage supply of the
emitter and replace If necessary
S013 Communication error
During important routines or when S062 occurred
30 times
Please contact SICK Customer
Service
S014 No result
No measval or ecorr file
S015
S016
S017
Motor x: defect
If, from system start, 30 step losses or 30 watchdog
actions
S018 Source failure
Detection I < 0.1 A
Check the emitter and replace If
necessary
S019 Chopper error
Detection: f_Motor < 50 Hz or sender SW signals
chopper error
Please contact SICK Customer
Service
S020 Configuration error
CONF (HC3X)
S021 Communication error
COM (HC3X)
S022 Controller not found
EXIST (HC3X)
S023 Frequent reset
Receiver, sender. If error occurred 30 times as from
system start
S024 No active component
If “Active” checkmarks of all components disabled Check in SOPAS ET