34
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
3
MAINTENANCE
3.4.6.1
Replacing the O-ring of the upper mirror plate
1
Remove the cell (
see Chapter “Removing the complete cell”, page 15
2
Loosen the fastening screws of the upper cover plate with a Phillips screwdriver
(4 pieces).
Fig. 37: Cell, fastening screws, cover plate 1
3
Loosen the fastening screw on the rear of the upper cover plate with a Phillips screw-
driver (1 piece).
Fig. 38: Cell, fastening screws, cover plate 1
4
Remove the upper cover plate.
5
Loosen the fastening screws on the upper mirror plate with a 5 mm Allen key (6 pieces).
1 Fastening screws of upper cover plate
Keep screws, spring washers and washers for reassembly.
1 Fastening screw, upper cover plate (rear side)
Keep screws, spring washers and washers for reassembly.