125
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
5.3.17
Replacing the PT100 temperature sensor
To replace the PT100, the cell and the sender/receiver unit must be completely removed.
1
Disconnect the MCS200HW power supply.
1
Remove the cell (
see Chapter “Removing the complete cell”, page 15
).
2
Remove the sender/receiver unit and the sheet metal covers (
the complete sender/receiver unit”, page 93
).
3
Loosen the cable conduit fastening clip with a 2.5 Allen key.
4
Remove the PT100 from the heating garland:
1
Open the contact plug.
▸
Carefully lever open the clamps of the contact plug with a slotted screwdriver
(3 pieces).
Fig. 195: Opening the contact plug
▸
Loosen the screws of the fastening clip inside with a slotted screwdriver (2 pieces).
Tools required
Phillips screwdriver
2 mm Allen key
2.5 mm Allen key
Spare parts set No.
Contents
2099813
PT100