5.7.11.12
Temperatures and Pressure
Specialist
: System edit: Temperatures and Pressure
The temperatures and the pressure correction managed by the system control are defined
in this menu option.
The heating controllers provide 12 bits (0 .. 4095 digits).
For default settings please refer to the status print-out in your system documentation.
Control function of heating controllers
The control function of the heating controllers is as follows:
• Temperature exceeding Hysteresis: Heating off
• Temperature lower than Hysteresis: Heating on
• Temperature lower than Actual Limit low:
Relay “ERROR” is set.
• If the temperature again exceeds this value, then:
Relay “ERROR” is reset.
• Temperature exceeding Actual value+Limit high:
Relays “ERROR” and ALARM are set.
• If the temperature again falls below this value, then:
Relay “ERROR” is reset; the relay “ALARM”, however, remains set. It must be reset by a
RESET.
The relays ERROR and ALARM are edited in Specialist: System edit: Zero/Calib/Purge
programs/other programs
.
For RESET of the alarms, the error message ER 32 (
Specialist: System edit: Zero/Calib/
Purge programs/other programs
is used.
Figure 20
Switching thresholds of the temperature controllers
Changing the temperature affects the measurement operation and can cause
instrument damage.
185 °C
197 °C
190 °C
180 °C
173 °C
Limit
Hysteresis
Nominal (=set value)
Hysteresis
Limit
Heating off
ERROR and ALARM set
ERROR reset
Heating on
ERROR set
ERROR reset
Example: Nominal = 185 °C
Hysteresis = 5 °C
Limit = 12 °C
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...