5.7.11
System edit (Name)
Specialist
: System edit
This menu option contains the configuration for the
system
(= the measurement sequence)
name
. The
system
to be configured can be:
• The actually active
system
. In this case any modifications to the settings have an
immediate effect on the configuration.
• A
system
that is not active at the moment (Selected in:
Specialist: System, select for
editing).
In this case, there is no effect on the actually active
system
. The
system
edited
can subsequently be activated via
Specialist: System, activate for measurement.
5.7.11.1
Signal output
Specialist
: System edit: Signal output
The following are defined here:
• Analog signal outputs
• Measurement ranges
Max. 32 outputs can be activated.
The menu option is identical to
Main Menu: Configuration: Signal output
Settings
Channel
Number of the analog output.
Mark (
)
Analog output active (
)/not active.
Name
Name of component derived from a list which is based on
Specialist: System edit:
Results
.
Display list by pressing <ENTER>
Unit
Unit of component. Displayed automatically from
Specialist: System edit: Results
Measurement
ranges
Switching of the measurement range is programmed here.
Measurement range 2 is the larger measurement range and relevant for the
plausibility check for zero setting- and for calibration, with the drifts etc.
The measurement ranges are automatically switched over (this is also true for the
graphic output). Switching over can be interrogated in the menu
Specialist: System
edit: Zero/Calib/Purge programs/other programs
(Variable RHi).
Hysteresis is 5%.
If one component has been defined more than once, then the 1st definition is
relevant for the plausibility check.
Relays
Number: Number of the relay
The relay “Number” is activated (deactivated) when the measurement range 2 (1) is
active (the relay indicates that the analog output switched to another output range).
0: No relay to be assigned
- Number: Inverted setting
Current
0..20 mA
4..20 mA
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...