5.7.9.3
Graphic output
Main Menu: Configuration: Graphic output
This menu option is used to define the graphic.
32 channels can be defined.
In
Time History
, 16 lines can be represented simultaneously, the first 8 lines with
corresponding values.
With the
bar graphs
, 4 (four) bars (of max. 16) can be represented simultaneously,
whereby it is possible to scroll with the arrow keys right/left.
The menu option is identical to
Specialist: System edit: Graphic output
5.7.9.4
Messages and corresp. relays
Main Menu: Configuration: Messages and corresp. relays
In this menu the relays that are marked with “
x
” in the upper line can be assigned to texts.
This means, the relay is set/reset corresponding to the text which is displayed on the
screen.
The free relays and texts are defined in the menu
Specialist: System edit: Messages and
corresp. relays
Settings
Channel
Numbering (no further significance)
Mark (
)
The channel is/is not displayed in the graphic
Name
Name of component Derived from a list which is based on
Specialist: System edit:
Results
. Display of list by <ENTER>.
Unit
Unit of component. Displayed automatically.
Limit values
When entering a mark (
), the limits defined in
Specialist: System edit: Limits
are
displayed as lines in the graphics.
Measurement
ranges
Scale of y-axis in time history.
2 scales are possible, which can be switched over automatically depending on the
measurement range.
Decimal
Defines the display of the measured values in the bar graph menu, e.g. xxxx.xxx = 3
digits following the decimal point. Adjustable by <ENTER>. In the time graphic, the
display of the numerical values shows always the complete sequence of digits
following the decimal point.
Key assignment:
F7
Copy line from signal output.
F8
Set standard (adopt measurement ranges from
Specialist: System edit: Signal
output
).
Settings
Relays
2 relays (left and right column) are possible.
Number
Number Number of the relay
(Selection only
possible, if marked with x in a.m. list)
0
No relay to be assigned
-Number Inverted setting
System status
The texts are displayed in the same order as defined in
Specialist: System edit:
Message and corresp. relays
.
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...