Maintenance
110
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
7.8.2
Watchdog for Communication
This watchdog controls the internal data communication.
Timeout: 4 minutes.
This watchdog cannot be disabled.
By removing a jumper it is possible to stop enabling the reset to the CPU.
Procedure
• Put the MCS100E out of operation (
) and disconnect from the power
supply.
• Open the MCS100E.
• Set the jumper of the watchdog:
Figure 22
Position and significance of jumper
• Close the MCS100E.
• Put the MCS100E into operation again (
WARNING:
Hot surface - risk of burns
The surface of the cell insulation is hot.
⊗
Do not touch the surface of the cell insulation.
Jumper
.
.
Jumper to the left: Reset of Watchdog enabled
Jumper removed: Reset of Watchdog disabled
Before the new start of the measurement sequence program, set the jumper to
the “left” again.