TECHNICAL SPECIFICATIONS
10
119
8020065-ZLP4/2017-08-10|SICK
Subject to change without notice
O P E R A T I N G I N S T R U C T I O N S | Master Data Analyzer
10.4
Data interface
The table below describes the structure of data emission via the data interface.
Column Field
Bytes
Description
Example
Note
1
Start of text
(special
character)
1
STX
2
Object number
(ON)
4
Measurement
identification index
0145
Internal index or external tray index
(optional).
Separator
1
;
;
3
Item
identification
(BC)
Varia
ble
Bar code content
of
one bar code
1JVGL0012345 May contain one bar code usually.
Also filtering functions to distinguish relevant
bar codes are available.
If no bar code could be read: NoRead is sent
out.
Separator
1
;
;
4
Weight
(OWE)
9
Gross weight from
scale
000081212
Weight value is zeroed if the measurement
was not legal for trade/ if there was an error.
Separator
1
;
;
5
Unit weight
(OWU)
1
Units of weight
g
Separator
1
;
;
6
Scale error
(connection)
(Byte 1 of SEID)
1
Scale is
connected/discon
nected
1
First byte of SEID:
0 for scale is connected.
1 for scale is not connected (e.g., not
powered/switched on).
Separator
1
;
;
7
Length
(OLE)
4
VMS measurement 0321
Zeroed if measurement error.
Separator
1
;
;
8
Width
(OLI)
4
VMS measurement 0150
Zeroed if measurement error.
Separator
1
;
;
9
Height
(OHE)
4
VMS measurement 0087
Zeroed if measurement error.
Separator
1
;
;
10
UnitVMS
(OUD)
2
Units of MVS
measurements
mm
mm
Separator
1
;
;
11
VMS status 1
(OMS1)
4
0000
0000 nor error
0001 item
outside
reading field
Hex. Format.
Output will be suppressed in case of an error.
Separator
1
;
;
12
Cubic volume
7
VMS measurement
LxWxH
mm3
Same as display output.
13
End of text
(special
character)
1
ETX
Terminator.
Tab. 31: Description of the data interface