OPERATION
Configuration
1. Log in to expert mode, see
2.
Define the parameter in the EXPRT-Config-MaskZn menu.
Note:
This setting can only be used in Pulse mode.
8.4.5
Selecting evaluation process
You can switch between Pulse mode and Foam mode for the evaluation process.
Different evaluation algorithms are used depending on the mode selected.
Configuration
1. Log in to expert mode, see
2.
Select either Pulse or Foam in the EXPRT-Config-Mode menu.
Where the following applies:
•
Mode = Pulse: The sensor measures either with or without AutCal.
•
Mode = Foam: The sensor measures only with valid CalMed. If there are
no valid calibrations, the message CalPIs is shown and the sensor goes into the
safe state.
Note:
If AutCal is accessed while the sensor is in Foam mode, AutCal is denied with the error
message !Denid.
8.4.6
Testing the configuration
Testing outputs
Switching / analog outputs can be simulated. This allows for the wiring and signal
values on the connected systems, such as the PLC control, relay, and lamps, to be
checked.
Configuration
Activate the Qx switching output
b
Set the parameter in the QxMENU-SimQx menu to QxOn.
Further options:
•
QxOff = switching output off
•
QxNorm = switching output in measuring operation
•
QxOn = switching output is active
Note:
The simulation is automatically deactivated if the supply voltage is interrupted.