MAINTENANCE
6
Replacing the sender lamp
1
Loosen both screws (5 mm Allen screw) on the sender lamp and take the sender lamp
off,
2
Remove the cap from the new sender lamp.
3
Plug in new sender lamp and screw tight.
4
Connect the plug and screw tight.
5
Plug in the lamp cover.
6
Screw the rear cover tight.
Adjustment is not required.
6.8
Cleaning the purge air unit
The filter of the purge air unit must be exchanged at the latest when the low-pressure
monitor on the filter outlet triggers.
Preparation
▸
If the purge air unit is not immediately functional: Remove the SR-unit from the gas duct
(swiveling out is sufficient for short-time work).
Procedure
1
Put the purge air unit out of operation and completely remove the purge air hoses.
2
Replace the air filter in the purge air unit and clean the inside of the purge air unit.
3
Fully swivel out the SR-unit so that any dust blown through the purge air hose is not
deposited on the window.
4
Put the purge air unit back into operation,
see “For GMP probe: Start-up of the purge air
.
NOTE:
Replacement of the sender lamp is identical for all GM32 variants.
NOTE: Inadequate purge air supply can damage the gas analyzer.
▸
The purge air unit must be in perfect condition.
Details
→
Data Sheet of purge air unit.