8012707/ZVF9/V2-1/2019-04 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32
Subject to change without notice
5
OPERATION
5.2.5.1
Diagnosis
The “Diagnosis” menu shows the current error messages.
Status messages, status indicators and system states
The GM32 signals a malfunction or an unsafe operational state with status signals
(option) (
→
electrical connection diagram).
The GM32 creates a logbook.
▸
Access to the logbook is performed exclusively via SOPAS ET
see “SOPAS ET (PC program)”, page 11
List of error messages and measures for clearing malfunctions
Status
Status
indicators
(LED)
Significance
Measured
value
display
Analog
outputs
[1]
Status
signal
[2],[3]
Power On
Green
Power supply in order
---
---
---
Uncertain
Green,
how-
ever, measured
value blinks
Measured value uncertain (e.g. outside
calibration range)
Cause: Press the
DIAG
button
All messages
→
SOPAS ET Logbook.
Clearing Malfunctions,
Current
Current
According to
setting
Mainte-
nance
request
Yellow
Irregularities (e.g. gas temperature too
high, deviation from check cycle too
high) that require a review of the cause.
The measured values are valid.
Cause: Press the
DIAG
button
All messages
→
SOPAS ET Logbook.
Clearing Malfunctions,
Current
Current
According to
setting
Failure
Red
Device failure (e.g. lamp failed)
Cause: Press the
DIAG
button
All messages
→
SOPAS ET Logbook.
Clearing Malfunctions,
Last valid
measured
value held
Last valid
measured
value held
According to
setting
Table 10
Status messages, status indicators and system states
[1] Option
[2] Option. Refer to the delivered System Documentation for the assignment of the status outputs.
[3] See SOPAS ET in the “Digital outputs” menu.